Pro-NM Performance

08-101 - standard version
08-103 - version with the PET Lid
Pro-NM Performance

The phantom for NM and PET systems performance evaluation (collimator, artifacts, calibration, reconstruction parameters). It can be used to evaluate, for example: center-of-rotation error, non-uniformity artifacts, changes of radius-of-rotation on spatial resolution, reconstruction filters on spatial resolution, attenuation and scatter compensation.


Technical data (can be modified to customer specifications):

  • main cylinder:
    • inside cylinder diameter: 206 mm
    • inside cylinder height: 186 mm
    • cylinder wall thickness: 7 mm
  • cold rods insert:
    • rod diameters: 4.8, 6.4, 7.9, 9.5, 11.1 and 12.7 mm
    • height of rods: 88 mm
  • cold spheres:
    • solid sphere diameters: 9.5, 12.7, 15.9, 19.1, 25.4 and 31.8 mm
    • height of the center of the spheres from the base plate: 127 mm
  • optional PET Lid with cylindrical samples (08-102):
    • refillable thin-walled cylinders, diameters: 8, 12, 16 and 25 mm
    • water filled cylinder diameter: 25 mm
    • air filled cylinder diameter: 25 mm
    • PTFE solid cylinder diameter: 25 mm
    • cylinder height: 38 mm
  • optional heavy duty carrying case (08-110)


Product features:

  • Complies with:
    • NEMA Standards Publication (NU 1-2001) Performance Measurements of Scintillation Cameras
    • AAPM Report No. 9 - Computer Aided Scintillation Camera Acceptance Testing
    • AAPM Report No. 22 - Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control
    • ACR–SNM  (Res. 5 – 2011) technical standard for diagnostic procedures using radiopharmaceuticals
  • CE certified
  • the Manual provides detailed guidelines for carrying out each test, results assessment and registration

Send an email. All fields with an * are required.

Name*:

E-mail*:

Phone*:

Firm*:

Country*:

Subject*:

Message*:

Send copy to yourself


Captcha verification*:
kod